RP-HPLC determination of meloxicam and its related substances in the meloxicam injection
作者
Yang Li
摘要
Objective To develope a specific,sensitive and rapid reversed phase HPLC method so as to determine meloxicam and its related substances in the meloxicam injection. Methods A kromsil—200 C 18 ( 200 mm× 4 6 mm , 5 μm ) column was used,the mobile phase was composed of methanol acetate buffer(56∶44, V∶V) and the UV detector was set at 270 nm. Results The linear calibration curves were obtained in the concentration range of 10~ 300 mg/L ( r =0 999 9 ). The mean recovery was 100 3%,RSD=1 2%. Conclusion This method is simple,rapid,accurate and suitable for the determination of meloxicam and its related substances in the meloxicam injection.