单室膝关节置换术
医学
外科
关节置换术
膝关节置换术
聚乙烯
口腔正畸科
材料科学
骨关节炎
复合材料
病理
替代医学
作者
V.N. Psychoyios,Ross Crawford,John O’Connor,David W. Murray
出处
期刊:The journal of bone and joint surgery
[British Editorial Society of Bone and Joint Surgery]
日期:1998-11-01
卷期号:80 (6): 976-982
被引量:178
标识
DOI:10.1302/0301-620x.80b6.8176
摘要
Many designs of unicompartmental knee replacement show early and mid-term failure due to polyethylene wear. We studied the wear rate of congruent polyethylene meniscal bearings retrieved from failed Oxford unicompartmental knee replacements. We examined 16 bearings, 0.8 to 12.8 years after implantation, measuring their thickness and comparing it with that of 14 unused bearings. The mean rate of penetration, which included the effects of wear at both upper and lower surfaces, was 0.036 mm per year (maximum 0.08). Bearings as thin as 3.5 mm wore no faster than thicker models, but ten with evidence of impingement had greater wear. The six bearings with no impingement showed a mean rate of penetration of 0.01 mm per year. In unicompartmental knee replacement, careful implantation of fully congruous meniscal bearings can avoid failure due to polyethylene wear.