Viviane Almeida Sarmento,Thais Feitosa Leitão de Oliveira Gonzalez,Ricardo Tadeu Lopes,Marco Novaes,Sirlene Mendes Borges,Izabel Regina Fischer Rubira‐Bullen
标识
DOI:10.1097/rct.0000000000001191
摘要
Objective The aim of this study was to evaluate the accuracy of multislice computed tomography (MSCT) and cone beam computed tomography (CBCT) scanners in the assessment of trabecular bone microarchitecture by radiomorphometric analysis. Methods Eight dry human hemimandibles were selected and scanned by 4- and 128-MSCT, CBCT with different voxel sizes, and a computed microtomography (micro-CT). The images were spatially aligned, the axial slices were segmented using CTAn software (Skyscan, Konitch, Belgium), and 10 bone morphometric parameters were quantified. The micro-CT images were the criterion standard. Results The results show the superiority of CBCT images with 0.2-mm voxel size and 128-MSCT. Conclusions We can conclude that the CBCT scanner with 0.2-mm voxel size and 128-MSCT had better performance on reproducibility of the distribution and organization of trabecular bone and marrow spaces evaluated in this study.