Niels R. van der Werf,Martin J. Willemink,Tineke P. Willems,Rozemarijn Vliegenthart,Marcel J. W. Greuter,Tim Leiner
标识
DOI:10.1007/s10554-017-1293-x
摘要
To evaluate the influence of heart rate on coronary calcium scores (CCS) using a dynamic phantom on four high-end computed tomography (CT) systems from different manufacturers. Artificial coronary arteries were moved in an anthropomorphic chest phantom at linear velocities, corresponding to < 60, 60-75 and > 75 beats per minute (bpm). Data was acquired with routinely used clinical protocols for CCS on four high-end CT systems (CT1-CT4). CCS, quantified as Agatston and mass scores were compared to reference scores at < 60 bpm. Influence of heart rate was assessed for each system with the cardiac motion susceptibility (CMS) Index. At increased heart rates (> 75 bpm), Agatston scores of the low mass calcification were similar to the reference score, while Agatston scores of the medium and high mass calcification increased significantly up to 50% for all CT systems. Threefold CMS increases at > 75 bpm in comparison with < 60 bpm were shown. For medium and high mass calcifications, significant differences in CMS between CT systems were found. Heart rate substantially influences CCS for high-end CT systems of four major manufacturers, but CT systems differ in motion susceptibility. Follow-up CCS CT scans should be acquired on the same CT system and protocol, and preferably with comparable heart rates.