Objective To establish a method for assay of homocysteine in plasma with high-performance liquid Chromatography.Methods plasma homocysteine was reduced by tributylphosphine and derivatized by fluorogenic thoil-specific reagent SBD-F.Derivatives of homocysteine were separated by fluorescence spectrophotometer.Results Calibration curve was linear in the range of 0~45 μmol/L for homocysteine.The recovery was 92.13%~108.04%.Intra-sample coefficient of variation was from 0.5 to 3.1%.Intra-and inter-day coefficients of variation were from 1.6% to 3.8% and from 2.8% to 4.6%.Conclusion The method that we have established has high sensitivity and good reproducibility and may be applied routine clinical use and scientific research.