Gonzalo Martínez‐García,Verónica Serafín,L. Agüı́,Paloma Yáñez‐Sedeño,José M. Pingarrón
标识
DOI:10.1002/elan.201400662
摘要
Abstract An electrochemical immunosensor for ghrelin (GHRL) determination in saliva is reported. Anti‐GHRL was immobilized onto Protein G‐magnetic beads and a competitive immunoassay involving biotinylated GHRL and alkaline phosphatase‐streptavidin was implemented. Once conjugate was magnetically captured on a screen‐printed carbon electrode, GHRL quantization was accomplished by DPV of 1‐naphtol formed upon addition of 1‐naphtyl phosphate. A linear range between 10 −3 and 10 3 ng/mL GHRL, and a LOD of 7 pg/mL, much smaller than those from commercial ELISA kits, were found. The usefulness of the immunosensor was demonstrated by analyzing human saliva spiked with GHRL at 0.01, 0.1, 1 and 10 ng/mL.