Rutherford backscattering spectroscopy (RBS) and low angle X-ray diffraction (LAXRD) were used to determine the density of thin Ta, Ag, Mo and Fe films as compared to the bulk value. An exact formula for the positions of the diffraction peaks as a function of the thickness of the film and its index of refraction was derived. A computer program utilizing this formula was written and used to do a least square fit to the low angle diffraction spectra and to obtain the thickness of the films. The densities of the films were found to be in the range between 84 and 98% of the bulk value. A detailed analysis of conditions that need to be fulfilled in order for the method to give reliable results is also presented.