Eran Leshem‐Rubinow,Cory M. Tschabrunn,Fernando M. Contreras‐Valdes,Israel Zilberman,Elad Anter
出处
期刊:Heart Rhythm
[Elsevier]
日期:2017-04-26
卷期号:14 (8): 1234-1240
被引量:27
标识
DOI:10.1016/j.hrthm.2017.04.035
摘要
Background An in vivo animal thigh model is the standard technique for evaluation of ablation catheter technologies, including efficacy and safety of ablation. However, the biophysics of ablation in a thigh model may not be similar to a beating heart. Objective The purpose of this study was to compare efficacy and safety of ablation between a thigh preparation model and a beating heart. Methods In 7 swine, radiofrequency ablation using a 3.5-mm open irrigated catheter (ThermoCool Smart Touch) was performed sequentially in a thigh muscle and in vivo beating ventricles. Ablation was performed at low (30 W for 40 s) and high (40 W for 60 s) energy settings and at similar contact force. Ablation lesions were scanned in high resolution and measured using electronic calipers. Results A total of 152 radiofrequency ablation lesions were measured (86 thigh and 66 heart). At low energy, lesion width was greater in the thigh model (12.19 ± 1.8 mm vs 8.99 ± 2.1 mm; P P = .18). The planar cross-sectional lesion area was greater in the thigh model (thigh 54.8 ± 10.8 mm 2 vs heart 43.1 ± 16.1 mm 2 ; P 2 vs heart 56.0 ± 15.5 mm 2 ; P Conclusion The thigh preparation model is a reasonable technique for evaluation of ablation catheter technology; however it often results in overestimation of lesion size, especially at higher energy settings.