钒
锰
铬
化学
吸附
吸附剂
柱色谱法
硅胶
石墨炉原子吸收
洗脱
原子吸收光谱法
检出限
吸附
核化学
无机化学
色谱法
量子力学
物理
有机化学
作者
Canan Ekinci Doğan,Ü. Köklü
标识
DOI:10.1080/10739140600609490
摘要
Abstract 3‐mercaptopropyltrimethoxysilane, (CH3O)3SiCH2CH2CH2SH, loaded on silica gel was used as a preconcentration sorbent for V, Cr, Mn and Pb prior to their determination by graphite furnace atomic absorption spetrometry (GFAAS). Surface characteristics and surface area of the silica gel before and after chemical modification were determined by elemental analysis. The retention and recovery of the analyte elements were studied by applying batch and column techniques. The experimental parameters, such as the effect of pH of the sample, shaking time in batch technique, flow rate of the eluent, the concentration of acid solution in the column section, and the amount of silica on retention and elution have been investigated. All elements were quantitatively (≥90%) recovered in the batch technique with R.S.D. values of 3.0 for vanadium, 1.8 for chromium, 1.7 for manganese, and 0.4 for lead. The same recoveries were obtained in the column techniques for all elements, while manganese in sea water could not be succesfully recovered. Detection limits of the method for vanadium, chromium, manganese, and lead are 1.1, 1.4, 1.3, and 0.8 ng, respectively.
科研通智能强力驱动
Strongly Powered by AbleSci AI