检出限
化学
杂质
色谱法
电子俘获探测器
碘化物
碘甲烷
分析化学(期刊)
气相色谱法
校准曲线
无机化学
有机化学
作者
Ajit Anerao,Bhushan Patil,R Jaguste,Nitin Pradhan
出处
期刊:Indian drugs
[Indian Drug Manufacturers' Association (IDMA)]
日期:2017-11-28
卷期号:54 (11): 39-45
被引量:1
标识
DOI:10.53879/id.54.11.11175
摘要
Methyl iodide has been highlighted as a potential genotoxic impurity (PGI). A sensitive HS-GC method with ECD was developed and validated for the determination of methyl iodide impurity in Tedizolid phosphate, an active pharmaceutical ingredient. HS-GC method on DB-624,column 60m X 0.53mm, film thickness 3μm, with electron capture detector (ECD) was used. The proposed method was specific, linear, accurate, rugged and precise. The calibration curves showed good linearity over the concentration range of 0.78 μg/g to 11.7 μg/g and the correlation coefficient was 0.997. Method had very low limit of detection (LOD) and limits of quantification (LOQ) are 0.23 μg/g and 0.78 μg/g, respectively, with respect to test concentration. Accuracy was observed within the range 96.0% to 103.2%. This method is a further extended good quality control tool for low level quantitation of methyl iodide impurity in another API, alogliptin benzoate.