血栓形成
闭塞
医学
心脏病学
生物医学工程
血小板
心肌梗塞
血小板活化
内科学
作者
Jess Berry,François J. Peaudecerf,Nicole A. Masters,Keith B. Neeves,Raymond E. Goldstein,Matthew T. Harper
出处
期刊:Lab on a Chip
[Royal Society of Chemistry]
日期:2021-01-01
卷期号:21 (21): 4104-4117
被引量:42
摘要
microfluidic assay. This highlighted the potential for occlusion to occur in thrombosis microfluidic devices through off-site coagulation, obscuring the effect of anti-platelet drugs. We therefore designed a novel occlusive thrombosis-on-a-chip microfluidic device that reliably generates occlusive thrombi at arterial shear rates by quenching downstream coagulation. We further validated our device and methods by using the approved anti-platelet drug, eptifibatide, recording a significant difference in the "time to occlude" in treated devices compared to control conditions. These results demonstrate that this device can be used to monitor the effect of antithrombotic drugs on time to occlude, and, for the first time, delivers this essential data in an unbiased and objective manner.