NIST公司
液体闪烁计数
放射化学
色谱法
化学
核工程
核物理学
物理
计算机科学
工程类
自然语言处理
作者
Karsten Kossert,Hendrik Bauer,Marcell Péter Takács,Ole Jens Nähle
标识
DOI:10.1007/s10967-025-10027-z
摘要
PTB has developed a new liquid scintillation counter for activity determination. The new system was designed to address the shortcomings of commercial devices when it comes to metrological scrutiny. It is equipped with three photomultiplier tubes and an automated sample changer. An external 133Ba gamma-ray source is used to determine two types of quench indicating parameters for the CIEMAT/NIST efficiency tracing (CNET) method. Simultaneously, the counting data are also available for TDCR analysis. In this paper, findings from several test measurements are presented to demonstrate the performance of the new counter for TDCR and CNET applications. The results show that the new counter is significantly superior to its commercial counterparts, particularly when determining quench indicators.
科研通智能强力驱动
Strongly Powered by AbleSci AI