光电阴极
光电流
检出限
化学
光电子学
干扰(通信)
纳米技术
分析物
材料科学
色谱法
计算机科学
电信
物理
频道(广播)
量子力学
电子
作者
Hanyu Wang,Dong Xue,Hui Wang,Xiang Ren,Huan Wang,Qin Wei,Dan Wu
出处
期刊:Talanta
[Elsevier BV]
日期:2023-05-11
卷期号:260: 124651-124651
被引量:3
标识
DOI:10.1016/j.talanta.2023.124651
摘要
In the realm of photoelectrochemical (PEC) analysis, it is customary to employ a lone photoanode or photocathode system for analyte detection. Nonetheless, such a single detection scheme inherently carries some deficiencies. While photoanode-based PEC immunoassay methods do exhibit conspicuous photocurrent responses and heightened sensitivity, they do suffer from inadequate resistance to interference when detecting in real sample detection. Photocathode-based analysis methods are capable of effectively surmounting the limitations of photoanode-based analysis methods, but their stability is poor. Based on the above reasons, this paper reports a novel immunosensing system combining ITO/WO3/Bi2S3 photoanode and ITO/CuInS2 photocathode. The system that incorporates both photoanode and photocathode exhibits a steady and discernible photocurrent, possesses robust resistance to external interference, and has effectively accomplished the quantification of NSE in the linear range from 5 pg/mL to 30 ng/mL. Remarkably, the detection limit has been determined to be 1.59 pg/mL. Besides the notable advantages of satisfactory stability, exceptional specificity, and outstanding reproducibility, the sensing system also introduces an innovative approach to fabricate PEC immunosensors.