Abstract A high‐performance liquid chromatography with ultraviolet detection method has been developed for the simultaneous determination of a set of reliable markers of renal function, including creatinine, uric acid, kynurenine and tryptophan in plasma. Separation was achieved by an Agilent HC‐C 18 (2) analytical column. Gradient elution and programmed wavelength detection allowed the method to be used to analyze these compounds by just one injection. The total run time was 25 min with all peaks of interest being eluted within 13 min. Good linear responses were found with correlation coefficient >0.999 for all analytes within the concentration range of the relevant levels. The recovery was: creatinine, 101 ± 1%; uric acid, 94.9 ± 3.7%; kynurenine, 100 ± 2%; and tryptophan, 92.6 ± 2.9%. Coefficients of variation within‐run and between‐run of all analytes were ≤2.4%. The limit of detection of the method was: creatinine, 0.1 µmol/L; uric acid, 0.05 µmol/L; kynurenine, 0.02 µmol/L; and tryptophan, 1 µmol/L. The developed method could be employed as a useful tool for the detection of chronic kidney disease, even at an early stage. Copyright © 2014 John Wiley & Sons, Ltd.