剂量计
再现性
材料科学
质量保证
线性
剂量分布
核医学
剂量学
医学
物理
数学
统计
量子力学
病理
外部质量评估
作者
S.W. Yang,M.J. Han,Dong Hoon Lee,S.K. Park
标识
DOI:10.1088/1748-0221/17/09/p09038
摘要
Abstract In this study, we developed a film-type perovskite dosimeter that can be used as a radiotherapy quality assurance (QA) dosimeter while replacing the existing semiconductor dosimeter. Dosimeter performance was evaluated by reproducibility, linearity, dose rate dependence, and percentage depth dose (PDD). In terms of reproducibility, the relative standard deviation (RSD) at 6 MV and 15 MV energy were 1.058% and 1.003%, respectively. As a result of linearity, a coefficient of determination of 0.9999 was presented at 6 MV and 15 MV energies. In terms of dose rate dependence evaluation, the RSD was 1.242% and 0.628% at 6 MV and 15 MV energies, respectively. As for PDD measurement, the result of matching the D max point was obtained. Our analysis confirms that the perovskite film dosimeter made of CsPbBr 3 was suitable for use as a QA dosimeter in radiotherapy.
科研通智能强力驱动
Strongly Powered by AbleSci AI