内科学
作者
Minoru Yamada,Yoshitake Yamada,Takehiro Nakahara,Shigeo Okuda,Takayuki Abe,Sachio Kuribayashi,Masahiro Jinzaki
标识
DOI:10.1016/j.jcct.2021.11.011
摘要
The development of ultra-high-resolution CT (U-HRCT) is expected to improve the accuracy of coronary stenosis evaluation. This study aimed to evaluate the accuracy of the stenosis severities of coronary artery phantoms estimated using U-HRCT by comparing them to those estimated with conventional CT.Coronary artery phantoms with non-calcified and calcified lesions were scanned with conventional CT (64-row × 0.625 mm) and U-HRCT (32-row × 0.3125 mm). The coronary artery phantoms had lumen diameters of 2.0, 3.0, and 4.0 mm with non-calcified lesions representing 0%, 25%, 50%, and 75% stenosis and 3.0 and 4.0 mm with calcified lesions representing 0%, 25%, 50%, and 75% stenosis. The lumen diameters at the stenotic and non-stenotic regions were measured, and the stenosis severities were compared with the true values.For non-calcified lesions, conventional CT significantly underestimated the stenosis severity in the phantom showing 75% stenosis with lumen diameters of 2.0 and 3.0 mm (p < 0.05), while the estimated stenosis severities were not significantly different from the true values at all settings with U-HRCT. For the calcified lesions, conventional CT overestimated the stenosis severities at all settings (p < 0.05), while U-HRCT yielded estimations closer to the true values, although still with some overestimation (p < 0.05).By using U-HRCT, the estimated stenosis severities of the coronary artery with non-calcified lesion become almost equal to the true value, while those with calcified lesion are still overestimated although they become closer to the true value.
科研通智能强力驱动
Strongly Powered by AbleSci AI