Tracing explosive in solvent using quantum cascade laser with pulsed electric discharge system
作者
Seong-Wook Park,Gang Chen,Anderson Chen,Chao Tian,Rainer Martini
出处
期刊:Applied Physics Letters [American Institute of Physics] 日期:2014-11-03卷期号:105 (18): 184104-184104被引量:1
标识
DOI:10.1063/1.4897440
摘要
We demonstrated highly sensitive detection of explosive dissolved in solvent with a portable spectroscopy system (Q-MACS) by tracing the explosive byproduct, N2O, in combination with a pulsed electric discharge system for safe explosive decomposition. Using Octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX), the gas was monitored and analyzed by Q-MACS and the presence of the dissolved explosive clearly detected. While HMX presence could be identified directly in the air above the solutions even without plasma, much better results were achieved under the decomposition. The experiment results give an estimated detection limit of 10 ppb, which corresponds to a 15 pg of HMX.