PURPOSE: To compare the parameters of three different topographic devices (Pentacam HR, Oculus Optikgeräte GmbH; Sirius, Costruzione Strumenti Oftalmici; and Cassini, i-Optics) in grading the severity of keratoconus in cross-linked and non-cross-linked eyes. METHODS: This was a prospective comparative interventional study done in a tertiary eye care center, wherein 114 eyes of 68 patients with keratoconus were divided into two groups: 62 eyes that were observed and 52 eyes that were cross-linked. All eyes were evaluated on all three topographers at baseline, 3-month follow-up, and 6-month follow-up. RESULTS: The Sirius showed significantly lower values of mean flat keratometry in comparison to the Pentacam HR (limits of agreement [LoA]: 1.75 to 3.51%) and Cassini (LoA: 1.75 to 3.51%). The mean steep keratometry values were higher for the Cassini in comparison to the Pentacam HR (LoA: 4.39 to 7.02%) and Sirius (LoA: 3.51 to 6.14%). The mean steep keratometry values of the Sirius were significantly lower than those of the Pentacam HR and in both the cross-linked and observation groups (LoA: 3.51 to 6.14%). The mean keratometry values were significantly higher in the Cassini compared to the Pentacam HR (LoA: 3.51 to 8.77%) and significantly lower in the Sirius in comparison to the Pentacam HR (LoA: 3.51 to 6.14%). The mean difference in astigmatism was also statistically significant between the three tomographers in both groups, with the Cassini showing higher values (LoA: 3.51 to 5.26%) and the Sirius lower values than the Pentacam HR (LoA: 5.26 to 6.14%). CONCLUSIONS: .