Ada S. Cheung,Casey de Rooy,Rudolf Hoermann,Emily Gianatti,Emma Hamilton,Geoff Roff,Jeffrey D. Zajac,Mathis Grossmann
摘要
Quantification of abdominal visceral adipose tissue (VAT) is important to understand obesity-related comorbidities. We hypothesized that dual X-ray absorptiometry (DXA) measurements of VAT would correlate with traditional gold standards of magnetic resonance imaging (MRI) and computed tomography (CT) in older men. Deming regression and Bland-Altman plots were used to assess the agreement between VAT measured simultaneously by DXA and MRI (n=95) in a cohort of older males participating in a randomized trial of testosterone replacement for diabetes. We also correlated DXA with single-slice CT (n=102) in a cohort of older males undergoing testosterone deprivation for prostate cancer. Lunar Prodigy DXA scanners using enCORE software was used to measure VAT. DXA VAT volume strongly correlated with MRI VAT volume (r=0.90, P<0.0001) and CT VAT area (r=0.83, P<0.0001). As DXA assesses VAT volume in a smaller compartment than MRI, Bland-Altman analysis demonstrated DXA systematically underestimated VAT by an approximately 30% proportional bias. DXA VAT volume measured by Lunar Prodigy DXA scanners correlate well with gold standard MRI and CT quantification methods, and provides a low radiation, efficient, cost-effective option. Future clinical studies examining the effects of interventions on body composition and regional fat distribution may find DXA an appropriate volumetric method to quantify VAT.