超声波传感器
收发机
波形
放大器
传感器
声学
压力传感器
传输(电信)
材料科学
电压
电子工程
计算机科学
工程类
电气工程
物理
CMOS芯片
机械工程
作者
Gerald Topalli,Yingying Fan,Matt Y. Cheung,Ashok Veeraraghavan,Mohammad Hirzallah,Taiyun Chi
标识
DOI:10.1109/tbcas.2024.3481414
摘要
This paper presents a 9-mW ultrasonic through-transmission transceiver (TRX) for portable, non-invasive intracranial pressure (ICP) sensing. It employs two ultrasound transducers placed at the temporal bone windows to measure changes in the ultrasonic time-of-flight (ToF), based on which the skull expansion and the corresponding ICP waveform are derived. Key components include a high-efficiency Class-DE power amplifier (PA) with 95% efficiency and an output swing of 15.8 VPP, along with a successive approximation register (SAR) delay-locked loop (DLL)-based time-to-digital converter (TDC) with 29.8 ps resolution and 122 ns range. Other than electrical characterization, the sensor is validated through two demonstrations using a water tank setup and a human head phantom setup, respectively. It demonstrates a high correlation of R
科研通智能强力驱动
Strongly Powered by AbleSci AI