Sihana Rugova,Marcus F. Abboud
出处
期刊:Bioengineering
[Multidisciplinary Digital Publishing Institute]
日期:2024-06-24
卷期号:11 (7): 642-642
被引量:10
标识
DOI:10.3390/bioengineering11070642
摘要
To ensure the prevention of thermal trauma and tissue necrosis during bone drilling in surgical procedures, it is crucial to maintain temperatures below the time- and temperature-dependent threshold of 50 °C for 30 s. However, the absence of a current standard for assessing temperatures attained during bone drilling poses a challenge when comparing findings across different studies. This article aims to address this issue by introducing a standardized testing method for acquiring thermal data during experimental bone drilling. The method requires the use of three controlled variables: infrared thermography, standard bone blocks, and a regulated drilling procedure involving a drill press with irrigation that simulates a surgeon. By utilizing this setup, we can obtain temperature data that can be effectively applied in the evaluation of other variables, such as surgical techniques or drill bit design, and translate the data into bone damage/clinical outcomes. Two surgical drill bits (2.0 mm-diameter twist drill bit and 3.3 mm-diameter multi-step drill bit) are compared using this experimental protocol. The results show the 2.0 mm bit reached significantly higher temperatures compared to the 3.3 mm bit when preparing an osteotomy (p < 0.05). The 2.0 mm drill bit reached temperatures over 100 °C while the 3.3 mm drill bit did not exceed 50 °C.