化学发光
单克隆抗体
免疫分析
材料科学
链霉亲和素
重复性
高尔基体
生物素
色谱法
纳米颗粒
磁性纳米粒子
化学
抗体
纳米技术
生物化学
生物
内质网
免疫学
作者
Xuefei Wei,Yanyan Xia,Mengjiao Shen,Yuan‐Han Yang,Jing Jin,Hongpan Xu,Zhiyang Li
标识
DOI:10.1166/jnn.2019.16485
摘要
Golgi protein 73 (GP73) is an independent diagnostic indicator of cirrhosis. However, it lacks automatic detection techniques to meet the large-scale clinical requirement in physical examination. In this paper, an automatic approach was established based on the ACL2800 automatic chemiluminescent analyzer, using magnetic nanoparticles (MNPs) and chemiluminescence. This method depended on sandwich strategy among biotin-labeled capture monoclonal antibody, target GP73 and acridinium ester (AE)-labeled reporter monoclonal antibody, conjugation of streptavidin-labeled MNPs to biotin-labeled capture monoclonal antibody, and chemiluminescent detection of AE-linked targets. Optimal conditions were investigated and clinical assessment was processed. Detection of GP73 demonstrated a high sensitivity of 1.19 ng/mL with a wide range from 1.34 ng/mL to 684.38 ng/mL. The quantitative detection was achieved with the repeatability of 2.69%, the coefficient of variation of 3.55% and the percent recovery of 93.46%-107.82%. Therefore, an automatic quantitative detection method was successfully developed, which was a potential screening method in physical examination.