Osvaldo Rampado,Paola Isoardi,R. Ropolo
标识
DOI:10.1088/0031-9155/51/6/015
摘要
Quality controls for testing the performance of computed radiography (CR) systems have been recommended by manufacturers and medical physicists' organizations. The purpose of this work was to develop a set of image processing tools for quantitative assessment of computed radiography quality control parameters. Automatic image analysis consisted in detecting phantom details, defining regions of interest and acquiring measurements. The tested performance characteristics included dark noise, uniformity, exposure calibration, linearity, low-contrast and spatial resolution, spatial accuracy, laser beam function and erasure thoroughness. CR devices from two major manufacturers were evaluated. We investigated several approaches to quantify the detector response uniformity. We developed methods to characterize the spatial accuracy and resolution properties across the entire image area, based on the Fourier analysis of the image of a fine wire mesh. The implemented methods were sensitive to local blurring and allowed us to detect a local distortion of 4% or greater in any part of an imaging plate. The obtained results showed that the developed image processing tools allow us to implement a quality control program for CR with short processing time and with absence of subjectivity in the evaluation of the parameters.