Fuli Li,Craig Kreikemeier‐Bower,Wanchuan Xie,Vishal Kothari,Benjamin S. Terry
摘要
A wireless medical capsule for measuring the contact pressure between a mobile capsule and the small intestine lumen was developed. Two pressure sensors were used to measure and differentiate the contact pressure and the small intestine intraluminal pressure. After in vitro tests of the capsule, it was surgically placed and tested in the proximal small intestine of a pig model. The capsule successfully gathered and transmitted the pressure data to a receiver outside the body. The measured pressure signals in the animal test were analyzed in the time and frequency domains, and a mathematic model was presented to describe the different factors influencing the contact pressure. A novel signal processing method was applied to isolate the contraction information from the contact pressure. The result shows that the measured contact pressure was 1.08 ± 0.08 kPa, and the small intestine contraction pressure's amplitude and rate were 0.29 ± 0.046 kPa and 12 min-1. Moreover, the amplitudes and rates of pressure from respiration and heartbeat were also estimated. The successful preliminary evaluation of this capsule implies that it could be used in further systematic investigation of small intestine contact pressure on a mobile capsule-shaped bolus.