脊髓损伤
医学
警报
计算机硬件
压力传感器
脊髓
生物医学工程
计算机科学
工程类
电气工程
机械工程
精神科
作者
Lingfeng Liu,Kang Zhao,Fan Chen,Jian Wu,Zhaofeng Yang,Ming Chen,Mao Lin-bo,Jiurong Han
摘要
Objective To develop a portable dynamic bladder pressure monitoring device for facilitating the monitoring of bladder detrusor pressure in patients with neurogenic bladder ( NB ) caused by spinal cord injury and further compared the effect with traditional urodynamics devices. Methods Portable dynamic bladder pressure monitoring device was implemented by software and hardware. The hardware was mainly composed of seven parts, such as micro liquid pressure sensor, electric bridge amplifying circuit, clock module, SD card storage circuit, liquid crystal display circuit, touch circuit, controller circuit and so on. The main program of the software included system initialization, data acquisition, storage, display, alarm, etc. A total of 45 patients with neurogenic bladder caused by spinal cord injury were enrolled in the study. Detrusor pressure was measured with the portable device and traditional urodynamics device. Results The test‐retest reliability of the portable device was modest in two times test (intra‐class correlation coefficient ( ICC ) = 0.852), no significant difference was found in the results of volume perfusion evaluation between our device and conventional urodynamic (each P > 0.05). Conclusion The portable dynamic bladder pressure monitoring device is conducive to the dynamic monitoring of bladder pressure in patients with neurogenic bladder caused by spinal cord injury.
科研通智能强力驱动