T. Fukuchi,Shin’ichiro Takeda,M. Katsuragawa,Goro Yabu,Shin Watanabe,Tadayuki Takahashi,Yuya Watanabe
标识
DOI:10.1088/1748-0221/18/01/p01030
摘要
Abstract We have developed a γ -ray computed tomography system using a CdTe double-sided strip detector. Owing to a 250 μm fine strip pitch and high energy resolution with photon-counting capability, the system provides highly accurate images, with which the materials and their distributions inside the target can be determined according to the photon transmittances. We evaluated the key performance of the system, conducting transmission measurements for Al, Cu, and Pb plates and also for Al, Fe, Cu, and Pb rod-phantoms, both using X-rays (∼30 keV) and γ -rays (∼80 keV) from a 133 Ba source. The measured transmittances agreed well with the calculated values from simulations. We successfully reconstructed the three-dimensional structure of the rod-phantom and distinguished the elements inside the phantom. Compared with the simulated photon transmittances, we found that material identification based on tomographic images obtained with the system is efficient as long as the target object does not contain thick high-Z elements.