克尔玛
乳腺摄影术
医学物理学
核医学
医学
放射科
核工程
剂量学
内科学
癌症
工程类
乳腺癌
作者
I Csete,L Büermann,Igor Gomola,Reinhard Girzikowsky
出处
期刊:Metrologia
[IOP Publishing]
日期:2013-01-01
卷期号:50 (1A): 06008-06008
被引量:5
标识
DOI:10.1088/0026-1394/50/1a/06008
摘要
A comparison of the air kerma standards for x-radiation qualities used in general diagnostic radiology and mammography, identified as EURAMET.RI(I)-S10 (EURAMET project #1221), was performed between the PTB and the IAEA. Two spherical and two parallel-plate reference-class ionization chambers of the IAEA and 12 beam qualities standardized in the IEC standard 61267:2005 plus 7 additional standard beam qualities established at both laboratories were selected for the comparison. The calibration coefficients were determined for the transfer chambers at the PTB in September 2012 and before and after this at the IAEA Dosimetry Laboratory. The results show the calibration coefficients of both laboratories to be in good agreement within the standard uncertainty of the comparison of about 0.47%. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by EURAMET, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
科研通智能强力驱动
Strongly Powered by AbleSci AI