胶囊
管(容器)
机器人
计算机科学
材料科学
地质学
工程类
人工智能
机械工程
古生物学
作者
Jinyang Gao,Zenglei Zhang,Guozheng Yan
出处
期刊:IEEE-ASME Transactions on Mechatronics
[Institute of Electrical and Electronics Engineers]
日期:2020-01-01
卷期号:: 1-1
被引量:22
标识
DOI:10.1109/tmech.2020.2999959
摘要
A clamper-based capsule robot (CCR) capable of exploring a compliant tube is promising for a minimally invasive medical application. It is composed of a locomotion unit and some functional modules (e.g., camera and telemetry circuit) at both ends; and the locomotion unit consists of a linear mechanism, a fixed clamper, and a free clamper that can be linearly moved by the linear mechanism. For the CCR design and optimization, analyzing the interaction between the CCR and the compliant tube is necessary. For this reason, this article conducts a quantitative locomotion analysis of the CCR in the compliant tube. Based on a minimum displacement assumption and by analyzing the tube deformation induced by each CCR locomotion gait, a model correlating the CCR locomotion efficiency with it design parameters was built. Then, the correctness of the model was validated in an intestine-constructed compliant tube. With the validated model, the influence of the design parameters on the CCR locomotion efficiency was analyzed to derive the design rules. In addition, a quantitative comparison between the locomotion efficiencies of the CCR and a previously developed inchworm-like capsule robot was conducted, showing that the CCR can enable a more efficient exploration for the compliant tube.
科研通智能强力驱动
Strongly Powered by AbleSci AI