万古霉素
治疗药物监测
医学
抗生素
全血
药品
药理学
重症监护医学
内科学
微生物学
细菌
生物
金黄色葡萄球菌
遗传学
作者
Ganesh S. Moorthy,Kevin J. Downes,Christina Vedar,Athena F. Zuppa
出处
期刊:Bioanalysis
[Future Science Ltd]
日期:2020-09-01
卷期号:12 (18): 1295-1310
被引量:17
标识
DOI:10.4155/bio-2020-0112
摘要
Background: Vancomycin is a commonly used antibiotic, which requires therapeutic drug monitoring to ensure optimal treatment. Microsampling assays are attractive tools for pediatric clinical research and therapeutic drug monitoring. Results: A LC–MS/MS method for the quantification of vancomycin in human whole blood employing volumetric absorptive microsampling (VAMS®) devices (20 μl) was developed and validated. Vancomycin was stable in human whole blood VAMS under assay conditions. Stability for vancomycin was established for at least 160 days as dried microsamples at -78°C. Conclusion: This method is currently being utilized for the quantitation of vancomycin in whole blood VAMS for an ongoing pediatric clinical study and representative clinical data are reported.