左乙拉西坦
化学
拉莫三嗪
色谱法
蛋白质沉淀
分析物
电喷雾电离
治疗药物监测
选择性反应监测
药品
质谱法
药理学
癫痫
串联质谱法
医学
精神科
作者
Shan Zhou,Ruitong Li,Ziyi Chen,Ruina Ren,Xueding Wang,Qilin Dai,Dingshen Wen,Yanping Guan,Xiaoxu Zhang,Siyuan Tang,Liemin Zhou,Min Huang
摘要
Abstract Background To minimize drug‐related toxicity and monitor dosing regimens, an ultra‐sensitive, simple and high‐throughput analytical method for therapeutic drug monitoring is required. A novel LC–MS/MS bioassay of levetiracetam, lamotrigine and 10‐hydroxycarbazepine in human plasma was established. The analytes were separated on a Hypersil GOLD™ C18 column under a 2.5 min isocratic elution after one‐step protein precipitation. MS detection was performed under electrospray ionization positive‐mode fitted with selected reaction monitoring. The validated ranges were 0.1–20 μg/ml for LTG, 0.3–60 μg/ml for 10‐hydroxycarbazepine and levetiracetam. The intra‐ and inter‐batches of precision and accuracy was within ±15%. The novel method met all other criteria. Conclusion: This method can be used to monitor drug concentrations and decision‐making in epileptic patients.
科研通智能强力驱动
Strongly Powered by AbleSci AI