摘要
With the increasing use of radiation therapy for treatment of canine sinonasal neoplasia, there is a need for developing low-dose CT protocols to help minimize radiation exposure. The purpose of this study was to assess the trade-off between image quality and reduced radiation exposure of a low-dose CT technique in the canine sinonasal cavity. In this prospective, experimental study, CT images of the sinonasal cavities from 10 normal Beagles were acquired using high-dose (130 kVp) or low-dose (110 kVp, 80 kVp) protocol. Radiation dose and image quality were compared. Radiation exposure measured by the volume-weighted CT dose index and dose-length product was reduced by 36% at 110 kVp and 74% at 80 kVp respectively, compared to the corresponding values at 130 kVp (P = 0.000). Low-dose protocol resulted in higher image noise and reduced signal-to-noise ratio and contrast-to-noise ratio than 130 kVp in most evaluated regions of interest (P < 0.05). CT numbers of the contrast-enhanced structures were highest at 80 kVp (P = 0.000). Conspicuity of most sinonasal structures was similar for high dose and both lower dose protocols. The results of this study indicate that 80 or 110 kVp can be used for sinonasal CT examinations to reduce radiation exposure to the patient without compromising image quality.