T. Jamneala,David A. Feld,D. Blackham,Ken Wong,B. Zaini
标识
DOI:10.1109/rfic.2006.1651192
摘要
Accurate characterization of the calibration standards used for the short-open-load-thru (SOLT) calibration of S-parameter measurements of RF devices is crucial if correct measurements are to be obtained. For most reciprocal devices (i.e. with S/sub 12/=S/sub 21/) if the calibration coefficients are inaccurate, the 'calibrated' S/sub 21A/ and S/sub 12A/ values appear not to be identical over all frequencies. We find that we can use this apparent violation of reciprocity to develop a novel procedure for standard characterization. This procedure ("reciprocal procedure") searches for the optimum set of calibration coefficients for which the reciprocity of the device is obeyed at all frequencies.