血糖自我监测
连续血糖监测
医学
血糖监测
静脉血
多中心研究
糖尿病
外科
内科学
随机对照试验
1型糖尿病
内分泌学
作者
Linong Ji,Lixin Guo,Junqing Zhang,Yufeng Li,Zhiyan Chen
标识
DOI:10.1177/19322968211037991
摘要
This study reports a clinical evaluation of AiDEX CGM system featuring a 14-day sensor, real-time glucose monitoring and factory-calibration. A multicenter, prospective, masked clinical study was conducted at with a total of 120 participants. Each participant wore 4 studied sensors and had one in-clinic visit for venous blood reference tests. 40 out of the 120 participants wore additional Abbott Libre sensors and performed at least 7 capillary BG tests daily for additional reference and comparison. Continuous glucose error grid analysis (CG-EGA) showed that AiDEX and Abbott Libre had good agreement with venous blood glucose, with 98.69% and 98.96% accurate readings, respectively. Overall MARD of AiDEX CGM systems was 9.08% when compared to venous blood reference and 10.1% when compared to finger capillary BG reference.