Objective To evaluate morphologic abnormalities warnings of the system XE-2100 hematology analyzer.Methods Manual stained differentiations were performed on 200 morphologic normal samples and 200 morphologic abnormal samples.Results The analyzer test showed the sensitivity to morphologic abnormalities was 99.34%,the specificity was 80.24%,the positive predictive value was 75.50% and the negative predictive value was 99.50%.The sensitivity to immature granulocytes was 95.31%,the specificity was 78.27%,the positive predictive value was 45.52% and the negative predictive value was 98.87%.The sensitivity to variant lymphocyte was 63.64%,the specificity was 96.47%,the positive predictive value was 83.58% and the negative predictive value was 90.39%.Conclusion The sample is warned by the system XE-2100 hematology analyzer must be manually differentiated.