Filippo Mele,Massimo Gandola,G. Bertuccio
标识
DOI:10.1109/tns.2021.3055934
摘要
A fully integrated pulsed reset charge-sensitive amplifier (CSA) optimized for low-capacitance (<; 100 fF) radiation detectors was developed on a 0.35-μm low-noise CMOS technology. The CSA was successfully tested with a collimated 10-mm 2 silicon drift detector (SDD) and extensively characterized. The experimental setup and the achieved spectrometry results with SDD are described for triangular and trapezoidal shaping filters with peaking times ranging from t peak =0.1-12.8 μs at an operating temperature of -35 °C. With a minimum equivalent noise charge (ENC) of 3.4 electrons rms measured at tpeak