Ultrathin EGD (UT-EGD) is an ideal tool for unsedated upper GI examination and pediatric gastroenterology but is rarely competent for EUS miniprobe (EUS-MP). We developed a UT-EGD US method (UT-EUS) and verified its clinical application value through animal experiments.
Methods
Five Bama miniature pigs were selected. Using an acoustic medium, we performed US on the duodenum, stomach, and esophagus, respectively, with conventional 20-MHz EUS miniprobe (EUS-MP-20), 20-MHz UT-EUS (UT-EUS-20), and 30-MHz UT-EUS (UT-EUS-30). The times to acquire 5 consecutive stable US images, number of identifiable wall layers, and quality and penetration depth of the images were recorded.
Results
No significant differences were found in the time required to obtain images between EUS-MP-20 and UT-EUS-20 at each site (P > .05). UT-EUS-30 showed more wall levels than UT-EUS-20 (P < .05). No significant differences were noted between EUS-MP-20 and UT-EUS-20 in imaging quality and penetration depth (P > .05).
Conclusions
The UT-EUS is easy to use with a satisfactory image quality and has potential clinical application value.