电化学发光
化学
检出限
电化学
电极
线性范围
无机化学
分析化学(期刊)
色谱法
物理化学
作者
Minli Yang,Chuanzhi Liu,Xinhui Hu,Pingang He,Yuzhi Fang
标识
DOI:10.1016/s0003-2670(02)00254-4
摘要
An electrochemiluminescence (ECL)-based method for rapid and sensitive detection of acridinium ester in neutral solution was described. Strong ECL emission was observed when a positive voltage over 2.0 V (versus Ag/AgCl) was applied to the working electrode (Pt) immersed in the acridinium ester solution of 2.0 mol l−1 KNO3 (pH 7.0). The possible ECL mechanism was discussed. It was proposed that the ECL emission came out of N-methylacridone, the oxidization product of acridinium ester by the nascent oxygen generated on the surface of working electrode in the course of oxidization of water. Other influenced factors including the electrochemical parameters, the ECL reaction medium and pH value, were investigated in detail. Under the optimal conditions, ECL intensity has a linear relationship with the concentration of acridinium ester in the range of 0.24–96 ng ml−1 (r=0.9999). The relative standard deviation for 24 ng ml−1 acridinium ester was 4.6% (n=11). The limit of detection was 0.16 ng ml−1.
科研通智能强力驱动
Strongly Powered by AbleSci AI