磁强计
灵敏度(控制系统)
光学
物理
电磁感应
无线电频率
激发
材料科学
电磁线圈
生物磁学
核磁共振
频域
医学影像学
度量(数据仓库)
声学
时域
电磁场
低频
迭代重建
磁场
电导率
图像传感器
频率调制
GSM演进的增强数据速率
作者
Cameron Deans,Luca Marmugi,Ferruccio Renzoni
摘要
Progress in electromagnetic induction imaging with atomic magnetometers has brought its domain to the edge of the regime useful for biomedical imaging. However, a demonstration of imaging below the required 1 Sm−1 level is still missing. In this Letter, we use an 87Rb radio frequency atomic magnetometer operating near room temperature in an unshielded environment to image calibrated solutions mimicking the electric conductivity of live tissues. By combining the recently introduced near-resonant imaging technique with a dual radio frequency coil excitation scheme, we image 5 ml of solutions down to 0.9 Sm−1. We measure a signal-to-noise ratio of 2.7 at 2 MHz for 0.9 Sm−1, increased up to 7.2 with offline averaging. Our work is an improvement of 50 times on previous imaging results and demonstrates the sensitivity and stability in unshielded environments required for imaging biological tissues, in particular for the human heart.