抗坏血酸
铜
检出限
流动注射分析
安培法
电极
化学
席夫碱
分析化学(期刊)
复合数
基础(拓扑)
色谱法
材料科学
核化学
复合材料
高分子化学
有机化学
数学
电化学
数学分析
物理化学
食品科学
作者
Mihajlo Kulizić,Milan Stanković,Marija Rašić,Aleksandar Mijatović,Rada Baošić,Aleksandar Lolić
标识
DOI:10.1002/slct.202304631
摘要
Abstract In this paper, a simple, sensitive, and precise flow injection system was optimized and applied to ascorbic acid determination with a carbon paste electrode modified with the Schiff base copper complex. The modified sensor showed increased sensitivity toward ascorbic acid oxidation with a decrease of oxidation potential by 200 mV. Under optimized flow parameters, working potential, gasket thickness, sample and reaction coil volumes, and flow rate, the developed flow system gave a linear response from 5 to 75 μmol/L of ascorbic acid. The calculated limit of detection was 4.5 μmol/L, calculated as 3σ/s which corresponds to the absolute value of 40 ng for the 50 μL sample coil. The precision for six consecutive injections of 5 and 75 μmol/L ascorbic acid solutions was determined as a relative standard deviation of 5.8 % and 1.5 %, respectively. The flow injection system enabled up to 120 analyses per hour. An optimized flow system was applied to the determination of ascorbic acid in dietary supplements, after simple preparation.
科研通智能强力驱动
Strongly Powered by AbleSci AI