生物医学中的光声成像
光声层析成像
带宽(计算)
断层摄影术
材料科学
光学
医学物理学
核医学
物理
计算机科学
医学
电信
作者
Geng Ku,Xueding Wang,George Stoica,Lihong V. Wang
标识
DOI:10.1088/0031-9155/49/7/018
摘要
Photoacoustic tomography, also referred to as optoacoustic tomography, employs short laser pulses to generate ultrasonic waves in biological tissues. The reconstructed images can be characterized by the convolution of the structure of samples, the laser pulse and the impulse response of the ultrasonic transducer used for detection. Although the laser-induced ultrasonic waves cover a wide spectral range, a single transducer can receive only part of the spectrum because of its limited bandwidth. To systematically analyse this problem, we constructed a photoacoustic tomographic system that uses multiple ultrasonic transducers simultaneously, each at a different central frequency. The photoacoustic images associated with the different transducers were compared and analysed. The system was tested by imaging both mouse brains and phantom samples. The vascular vessels in the brain were revealed by all of the transducers, but the image resolutions differed. The higher frequency detectors provided better image resolution while the lower frequency detectors delineated the major structural traits with a higher signal-noise ratio.