γ能谱仪
分光计
半导体探测器
脉搏(音乐)
γ射线光谱法
物理
计算机科学
核物理学
光学
放射化学
探测器
化学
作者
Hesham H. Shaker,Sofia Salama,Hany Kasban,A.A. Saleh,Mohamed Amin Dessouky
标识
DOI:10.1088/1748-0221/20/03/p03007
摘要
High-purity germanium (HPGe) detector is one of the best detectors for gamma spectrometry due to its excellent energy resolution. In a previous publication, we were interested in implementing a low-cost digital pulse processing (DPP) unit for sodium iodide detectors. This implementation was based on the SAM3X8E microcontroller unit. In this paper, we use the same microcontroller unit to develop a DPP unit for an HPGe detector. The developed unit is compared with a commercial device. The achieved energy resolution of the developed unit is worse than what is achieved by a commercial unit, but it still can be used as a low-cost alternative. The proposed DPP unit is ideal for identifying radioisotopes in applications prioritizing affordability, portability, and ease of use over high-end resolution. It is suitable for environmental monitoring, nuclear security, and educational labs, benefiting from its cost-effective and accessible design.
科研通智能强力驱动
Strongly Powered by AbleSci AI