Aria M. Salyapongse,Gregory David Avey,James Fink,Michael Howard Lev,Jeffrey R. Sachs,Zhye Yin,Manmohan Singh,Nicholas K. Mathew,Giuseppe V. Toia,Timothy P. Szczykutowicz
摘要
The purpose of this work was to evaluate image quality of non-contrast neurological CT scans on a commercially available energy-integrating detector (EID) CT compared to a prototype deep silicon (dSi) photon-counting detector CT (PCCT). A cohort of subjects scanned on the EID-CT (Apex CT, GE HealthCare, Waukesha, WI) was retrospectively recruited. A second cohort of prospectively recruited subjects was scanned on the prototype dSi-PCCT CT using deep learning (DL) recon. Image quality was assessed by four neuroradiologists on a 5-point Likert scale for sharpness, noise, and presence of artifacts, differentiation of gray-white matter, and structural confidence in the lentiform nucleus and insular ribbon. Reader agreement was assessed through a linearly weighted Cohen's kappa. Contrast-to-noise ratio (CNR) between gray and white matter was determined. Differences between the median Likert scores and CNRs of the two cohorts were assessed with the Mann-Whitney U test. Reader agreement was fair on average ([Formula: see text]). Median Likert scores of the dSi-PCCT with DL recon cohort were statistically superior compared to the median scores of the EID-CT scores for sharpness, noise, gray-white matter differentiation, and structural confidence of the lentiform nucleus and insular ribbon. No statistical difference was detected in the presence of artifacts between the cohorts. Average CNRs were 1.45±0.14 and 1.53±0.08 for the EID-CT and dSi-PCCT cohorts, respectively, but no significant difference was detected (p = 0.72). Increased sharpness, lower noise, and better ability to differentiate structures demonstrate improved image quality and the potential for utilizing PCCT in non-contrast neurological imaging tasks.
科研通智能强力驱动
Strongly Powered by AbleSci AI