声辐射力
激光多普勒测振仪
传感器
光学
声学
物理
多普勒效应
镜头(地质)
振动
焦距
光学(聚焦)
相(物质)
超声波
激光器
量子力学
分布反馈激光器
天文
作者
Matthew W. Urban,Miguel Bernal,James F. Greenleaf
出处
期刊:IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control
[Institute of Electrical and Electronics Engineers]
日期:2007-06-01
卷期号:54 (6): 1142-1153
被引量:23
标识
DOI:10.1109/tuffc.2007.368
摘要
We describe a phase aberration correction method that uses dynamic ultrasound radiation force to harmonically vibrate an object using amplitude modulated continuous wave ultrasound. The phase of each element of an annular array transducer is adjusted to maximize the radiation force and obtain optimal focus of the ultrasound beam. The maximization of the radiation force is performed by monitoring the velocity of scatterers in the focus region. We present theory that shows focal optimization with radiation force has a well-behaved cost function. Experimental validation is shown by correction of manual defocusing of an annular array as well as correcting for a lens-shaped aberrator placed near the transducer. A Doppler laser vibrometer and a pulse-echo Doppler ultrasound method were used to monitor the velocity of a sphere used as a target for the transducer. By maximizing the radiation force-induced vibration of scatterers in the focal region, the resolution of the ultrasound beam can be recovered after aberration defocusing.