腹部
核医学
磁共振成像
氘
电磁线圈
胃
核磁共振
体素
化学
放射科
医学
生物医学工程
内科学
物理
量子力学
电气工程
工程类
作者
Mary A. McLean,Ines Horvat‐Menih,Pascal Wodtke,Joshua Kaggie,Jonathan R. Birchall,Rolf F. Schulte,Ashley Grimmer,Elizabeth Latimer,Marta Wylot,Maria J Zamora-Morales,Alixander S. Khan,Huanjun Wang,James A. Armitage,Thomas J. Mitchell,Grant D. Stewart,Ferdia A. Gallagher
摘要
Abstract Purpose To establish and optimize abdominal deuterium MR spectroscopic imaging in conjunction with orally administered 2 H‐labeled molecules. Methods A flexible transmit‐receive surface coil was used to image naturally abundant deuterium signal in phantoms and healthy volunteers and after orally administered 2 H 2 O in a patient with a benign renal tumor (oncocytoma). Results Water and lipid peaks were fitted with high confidence from both unlocalized spectra and from voxels within the liver, kidney, and spleen on spectroscopic imaging. Artifacts were minimal, despite the high 2 H 2 O concentration in the stomach immediately after ingestion, which can be problematic with the use of a volume coil. Conclusion We have shown the feasibility of abdominal deuterium MR spectroscopic imaging at 3 T using a flexible surface coil. Water measurements were obtained in healthy volunteers, and images were acquired in a patient with a renal tumor after drinking 2 H 2 O. The limited depth penetration of the surface coil may have advantages in characterizing early uptake of orally administered agents in abdominal organs, despite the high concentrations in the stomach which can pose challenges with other coil combinations.
科研通智能强力驱动
Strongly Powered by AbleSci AI