化学
激光诱导击穿光谱
激光器
脉搏(音乐)
光谱学
分析化学(期刊)
光学
信号(编程语言)
原子物理学
物理
色谱法
量子力学
探测器
计算机科学
程序设计语言
作者
Zuhaib Haider,Yusof Munajat,Raja Ibrahim Kamarulzaman,Nurhidayu Shahami
标识
DOI:10.1080/00032719.2014.940532
摘要
A simple and cost-effective variant of laser induced breakdown spectroscopy is presented that involves a double simultaneous pulse configuration employing a single laser source. Its performance is compared with conventional single pulse configuration. Double simultaneous pulses were accomplished by splitting a Nd:YAG laser (1064 nm, 6 ns, 360 mJ) beam into two components that were focused on the sample surface to produce two concurrent breakdowns. Experiment was repeated for single pulse and double simultaneous pulses under different ambient pressures. The performance was evaluated on the basis of self-absorption, signal-to-noise ratio (SNR), and relative standard deviation (RSD) of the Mg II doublet (280.2704 nm, 279.553 nm). Optically thin emission lines of better profiles with higher signal-to-noise ratio resulted from double simultaneous pulses. The lowest relative standard deviations obtained by single pulse and double simultaneous pulse configurations were 18.89% and 12.01%, respectively. In fact, double simultaneous pulses have performed better than single pulse in all respects within the studied regime.
科研通智能强力驱动