颈椎前路椎间盘切除融合术
冯·米塞斯屈服准则
笼子
运动范围
植入
有限元法
固定(群体遗传学)
生物力学
材料科学
口腔正畸科
脊柱融合术
颈椎
颈椎
医学
结构工程
解剖
外科
工程类
环境卫生
人口
标识
DOI:10.1615/jlongtermeffmedimplants.2022044668
摘要
One of the standard treatments for spinal diseases is anterior cervical discectomy and fusion (ACDF). ACDF is a secure and successful operation that prevents patients to improve their pain and function. The mechanical goal of the ACDF is to prevent motion between adjoining vertebrae by a novel cage-screw implant. The objective of this study is to analyze the biomechanical flexibility in terms of the range of motion (ROM) of two-level ACDF fixation using the finite element method (FEM). A CT scan-based FEM model of the cervical spine (C2-C7) is used and two-level cage is implanted at C4-C6 segments. A 50-N compressive force and 1-Nm moment are applied on C2 vertebrae and C7 is fixed in all directions. The ROM at two-level fixation (C4-C5-C6) is reduced by 55 to 88% compared with intact spine during all physiological movement. The ROM slightly increase (3-9%) at the adjacent segment. The maximum von Mises stress variations are 25-65 MPa during flexion-extension, lateral bending, and axial rotations under given loading. The maximum von Mises stress found in cage and screw is below the yield stress during all physiological movement.
科研通智能强力驱动
Strongly Powered by AbleSci AI