Robby Vanspauwen,Floris L. Wuyts,Paul Van de Heyning
标识
DOI:10.3233/ves-2009-0358
摘要
OBJECTIVE: To determine normal limits and to analyse the test-retest reliability of the vestibular evoked myogenic potentials (VEMPs) parameters. METHODS: The VEMP procedure was repeated on different test days to analyze test-retest differences. We calculated several reliability parameters: intraclass reliability coefficient (ICC), method error (ME), coefficient of variation of the method error (CV(ME)), standard error of measurement (SEM) and minimal difference (MD) for test-retest measurements. Normal values for left-right differences, based on the interaural ratio (IAR), were determined. RESULTS: For each VEMP parameter, the ICC values indicated excellent reliability, except for p13 and corrected amplitude (fair to good reliability). The CV(ME)) values were less than 7% for p13, n23, threshold, MRV(females) and MRV(males). For the parameters corrected amplitude and raw amplitude, the CV(ME)) values exceeded 15%. The 95% IAR prediction intervals (PIs) were also largest for the parameters raw amplitude and corrected amplitude. CONCLUSION: In order to evaluate a VEMP outcome in a patient, the VEMP parameters and IAR values can be compared with the 95% PI of the normal values. When successive measurements are performed within the same subjects, the minimal difference (MD) serves as a tool to decide whether these differences are clinically relevant or not.