摘要
The in vitro cavitation thresholds and locations were studied on ten different heart valve designs. The valves were mounted in a mock circulation flow loop which simulated a cardiovascular system. All the tests were run at 70 beats per minute with a cardiac output varying between 2 l/min and 6 l/min in increments of 1 l/min. In vitro cavitation phenomena generated at the closing instant of mechanical heart valves were captured using a video photographic technique. Cavitation locations and intensity on different valve designs were analyzed from the cavitation images recorded on a video tape. When cavitation occurs on a bileaflet valve, it can occur in the same localized area of the leaflet from cycle to cycle thus producing a cumulative effect. In a single disc valve, the free rotation of the valve disc during operation provides a means of distributing a localized cavitation activity over an ever changing disc surface. Thus any cavitation-induced damage on the disc surface can be reduced or eliminated even though a single disc valve may have a lower cavitation threshold. Cavitation locations and thresholds are primarily a function of valve design. Smaller size valves have higher cavitation thresholds than larger ones. The cavitation thresholds of all the valves tested were above the physiological left ventricular maximum dp/dt at rest. If in vivo cavitation occurs under some extreme conditions, this study suggests possible locations on mechanical heart valves which could be examined for traces of cavitation activity.
科研通智能强力驱动
Strongly Powered by AbleSci AI