垂直的
材料科学
剪切(地质)
生物力学
剪切力
生物医学工程
口腔正畸科
解剖
复合材料
医学
数学
几何学
作者
Joan E. Sanders,Colin Daly,Ernest M. Burgess
标识
DOI:10.1682/jrrd.1992.10.0001
摘要
Shear stresses on a residual limb in a prosthetic socket are considered clinically to contribute to tissue breakdown in below-knee amputees. When applied simultaneously with normal stresses, they can cause injury within the skin or can generate an abrasion on the surface. To gain insight into shear stresses and parameters that affect them, interface stresses were recorded on below-knee amputee subjects during walking trials. On the tibial flares, resultant shear ranged from 5.6 kPa to 39.0 kPa, while on the posterior surface it ranged from 5.0 kPa to 40.7 kPa. During stance phase, anterior resultant shears on a socket were directed toward the apex while posterior resultant shears were directed downward approximately perpendicular to the ground. Waveform shapes were usually double-peaked, with the first peak at 25% to 40% into stance phase and the second peak at 65% to 85% into stance. Application of these results to residual limb tissue mechanics and prosthetic design is discussed.
科研通智能强力驱动
Strongly Powered by AbleSci AI