钛
植入
材料科学
梁(结构)
涂层
生物医学工程
骨整合
表面改性
髁突
医学
复合材料
外科
光学
冶金
化学
物理
物理化学
作者
Jacob Biemond,René Aquarius,Nico Verdonschot,Pieter Buma
标识
DOI:10.1007/s00402-010-1218-9
摘要
Electron beam melting (E-beam) is a new technology to produce 3-dimensional surface topographies for cementless orthopedic implants.The friction coefficients of two newly developed E-beam produced surface topographies were in vitro compared with sandblasted E-beam and titanium plasma sprayed controls. Bone ingrowth (direct bone-implant contact) was determined by implanting the samples in the femoral condyles of 6 goats for a period of 6 weeks.Friction coefficients of the new structures were comparable to the titanium plasma sprayed control. The direct bone-implant contact was 23.9 and 24.5% for the new surface structures. Bone-implant contact of the sandblasted and titanium plasma sprayed control was 18.2 and 25.5%, respectively.The frictional and bone ingrowth properties of the E-beam produced surface structures are similar to the plasma-sprayed control. However, since the maximal bone ingrowth had not been reached for the E-beam structures during the relatively short-term period, longer-term follow-up studies are needed to assess whether the E-beam structures lead to a better long-term performance than surfaces currently in use, such as titanium plasma spray coating.